A High Sensitive Chemiluminescence Analysis System For Medical Radiation Dose Evaluation

X Li,Yd Wei,D Xing,Yh He
DOI: https://doi.org/10.1117/12.482943
2002-01-01
Abstract:Chemiluminescence (CL) analysis technique is an important method in biological and medical diagnosis. In this research work, we present a novel system to detect and analyze the CL during the actinotheraphy to evaluate the radiation dose according to the principle of the cell stress response to ionizing radiation. The system consists of a sample chamber, a photomultiplier tube (PMT) detector, an amplifier, a discriminator, and a pulses counter. The PMT was cooled to reduce the thermal noise. With the single photon counting technique, high sensitivity, signal-to-noise ratio and stability was achieved. The sample chamber, which was driven by a step motor, has six sample cells for control experiments. A personal computer dominates the data acquisition and the control of the system through the newly implemented universal serial bus interface. There is also a peristaltic pump for auto-injecting the chemical reagent solution. The software developed on the LabVIEW platform makes the system with high automation. With the system, radiation dose could be quantified by the CL intensity from the cell stress response to ionizing radiation. Leukocytes, CCRF cell line and a CL probe MCLA were used to test the reliability of this system. The relationship between the (CO)-C-60 irradiation doses and the intensity of the MCLA CL were obtained, which showed that this system could image the CL properly.