Determination of Trace Metal Impurities in High Pure Tin Oxide Electrode Material by Inductively Coupled Plasma Mass Spectrometry
Xu Wei,Li Yu-Zhen,Duan Tai-Cheng,Chen Hang-Ting
DOI: https://doi.org/10.11895/j.issn.0253-3820.150577
IF: 1.193
2015-01-01
Chinese Journal of Analytical Chemistry
Abstract:A new method for determination of trace Cu, Cr, Mn, Co, Ni, Cd, Fe and Pb impurities in high pure tin oxide material was proposed. 0. 1 g of sample was successfully digested at 230 degrees C in 3 h when autoclaves with high pressure and high temperature were used. After that, over 99. 9% of the matrix Sn was removed by chloride volatilization at 150 degrees C for 1. 5 h, which eliminated the matrix effect and spectra interference in the following ICP. MS detection. The recoveries for all the analytes of the spiked test were 100%+/- 5%, the RSDs for 6 runs of sample analysis were from 0. 09% to 3. 68%, and the limits of detection (LOD, 3 sigma) were from 0. 002 mg/kg to 0. 034 mg/kg, which indicated that the established method were of both accuracy and high sensitivity, and could be applied for the analysis of samples with purity up to 6N. Finally, three samples specified with 4N purity were analyzed by this method.