Determination Of The Trace Amount Of 2-Diphenylamine And 4-Diphenylamine In Waste Water By Ion Chromatography With Fluorescence Detector

Yan Zhu,Yintu Jiang,Mingli Ye,Dan Cuo,Guanghong Xiang
DOI: https://doi.org/10.3321/j.issn:0253-3820.2001.09.009
IF: 1.193
2001-01-01
Chinese Journal of Analytical Chemistry
Abstract:2-Diphenylamine(2-ADP) and 4-diphenylamine(4-ADP) in water were detected by Dionex 4000i ion chromatograph with fluorescence detector. This experiment was performed by OmniPac PCX-500 column using 0.06 mol/L NaCl, 0.08 mol/L HCl, 0.40 V/V ACN as eluate. The fluorescence detector was measured in excitation and emission wavelengths of 230 mn 395 nm, respectively. The detection limit was 0.006 mg/L and 0.010 mg/L for 2-ADP and 4-ADP, respectively, with good linear relation, selection and sensitivity. The method has been applied to the determination of trace amount of 2-ADP and 4-ADP in waste water samples.
What problem does this paper attempt to address?