Calibrasion of the energy response for the soft x-ray detections elements with the beijing- synchrotron radiation facility
SUN KE-XU, YI RONG-QING, YANG JIA-MIN, WANG HONG-BIN, MA HONG-LIANG, CHEN ZHENG-LIN, HUANG TIAN-XUAN, CUI YAN-LI, ZHENG ZHI-JIAN, TANG DAO-YUAN, DING YONG-KUN, WEN SHU-HUAI, JIANG WEN-MIAN, ZHAO YONG-KUAN, CUI MING-QI, LI GANG, CUI CONG-WU, TANG E-SHENG
DOI: https://doi.org/10.7498/aps.46.650
1997-01-01
Abstract:Calibration of the energy response for the soft X-ray detection elements using in the ICF diagnostic are reported.The Beijing synchrotron Radiation Facility (3B1 beam line,photon energy~250—1000eV,radiation flux ~1012photon/s·mm2·mr2·0.1% band-width,beam current ~20—80mA) and target chamber of the reflectometer are used.The source intensity is monitored by AXUV-100 photodiode.The calibration of the energy response for the XRD,filter light film,Ⅱa type diamond photoconductive detector,transmission grating and soft X-ray film are given and analysed.