Imaging optimization of fluorescence confocal scanning system

Ping Zhang,Zhen Wu,Cuiying Wang,Jiying Xian
1997-01-01
Abstract:The mathematical model of 3-dimensional optical transfer function (OTF) was obtained by Fourier transform of point spread function for a fluorescence confocal scanning microcopy system. From this model, The 3-dimensional OTF was computed for different sized annular lenses and detectors. By using computer simulation and numerical computation of OTF, the image property was analyzed for the case of different annular lenses and detector in confocal scanning system.
What problem does this paper attempt to address?