Determination of arsenic species by hydride generation/gold trapping/gas chromatograph/atomic absorption spectrometry in natural water

Puhua Chen,Weiqi Chen,Jianhui Zhang,ShuGui Dai
1996-01-01
Abstract:A method of hydride generation/cold trapping/gas chromatography/atomic absorption combination has been set up. Chromosorb GAW-DMGS(particle size 0.3��0.45 mm) coated 3% OV-101 is used as chromatographic packing material. Detection limits in terms of As are As(V) 0.51 ng, As(III) 0.43 ng, MMA 0.38 ng, DMA 0.67 ng, and the relative standard deviations for 12 ng As are As (V) 4.21%, As (I) 3.56%, MMA 3.23%,DMA 5. 46%, respectively. Linearity of calibration curves in the range of 0��50 ng As is very good. This method was used to determine arsenic species in lake, river, sea and underground waters. Recoveries of each species in water sample are between 93.5% and 104.9%. The analytical results for the four arsenic species are given.
What problem does this paper attempt to address?