Near-Infrared Methane Detection Device Using Wavelength-Modulated Distributed Feedback Diode Laser Around 1.654 Μm

Jian-Qiang Huang,Chuan-Tao Zheng,Zong-Li Gao,Wei-Lin Ye,Yi-Ding Wang
DOI: https://doi.org/10.1080/00387010.2013.790062
2013-01-01
Spectroscopy Letters
Abstract:ABSTRACT By using a wavelength-modulated distributed feedback laser centered at 1.654 µm, a stand-alone near-infrared methane detection device was experimentally proposed based on the tunable diode laser absorption spectroscopy technique. An intelligent temperature controller, a scan and modulation module, and a cost-effective lock-in amplifier were developed to drive the distributed feedback laser and extract the second harmonic signal. Experimental results show that the relative detection error is less than 7% within the detection range of 0–106 ppm, and the limit of detection is about 11 ppm with an absorption length of 0.2 m. Long-term monitoring on two gas samples (103 ppm and 2 × 105 ppm) suggests good stability with the maximum detection errors smaller than 7% and 2.5%, respectively. Due to careful design and integration, the developed near-infrared sensor reveals competitive performances compared with our previously reported sensing devices at the mid-infrared region.
What problem does this paper attempt to address?