Correlations between b-mode ultrasound image texture features and tissue temperature in hyperthermia

Lei Sheng,ZHOU Zhu-huang,WU Shui-cai,Yi Zeng
2013-01-01
Abstract:The noninvasive estimation of tissue temperature is a key problem in hyperthermia. The original ultrasonic RF signal was used to reconstruct B-mode images, eliminating device-dependent variances to explore the use of B-mode image texture parameters in monitoring temperature and coagulation zones in hyperthermia. The water bath heating experiments were carried out on in-vitro porcine liver samples, and the ultrasound RF signal at different temperatures was collected. Then the B-mode ultrasonic images were reconstructed, and the texture features (gray level histogram, gray level co-occurrence matrices, and gray level-gradient co-occurrence matrix) were extracted. The correlations between temperature and texture parameters of B-mode images were analyzed. Results indicate that some texture parameters are linearly correlated with the temperature in the heated tissue at a temperature from 20°C to 60°C. Five parameters with large correlations with temperatures include mean gray scale of gray level histogram, entropy of gray level co-occurrence matrix, and hybrid entropy, inverse difference and correlation of gray level-gradient co-occurrence matrix. These five texture features with large correlations with temperatures can be used for tissue temperature estimation in clinical microwave tumor ablation.