Shape-recovery of implanted shape-memory devices remotely triggered via image-guided ultrasound heating
Yang Zhu,Kaicheng Deng,Jianwei Zhou,Chong Lai,Zuwei Ma,Hua Zhang,Jiazhen Pan,Liyin Shen,Matthew D. Bucknor,Eugene Ozhinsky,Seungil Kim,Guangjie Chen,Sang-ho Ye,Yue Zhang,Donghong Liu,Changyou Gao,Yonghua Xu,Huanan Wang,William R. Wagner
DOI: https://doi.org/10.1038/s41467-024-45437-2
IF: 16.6
2024-02-07
Nature Communications
Abstract:Shape-memory materials hold great potential to impart medical devices with functionalities useful during implantation, locomotion, drug delivery, and removal. However, their clinical translation is limited by a lack of non-invasive and precise methods to trigger and control the shape recovery, especially for devices implanted in deep tissues. In this study, the application of image-guided high-intensity focused ultrasound (HIFU) heating is tested. Magnetic resonance-guided HIFU triggered shape-recovery of a device made of polyurethane urea while monitoring its temperature by magnetic resonance thermometry. Deformation of the polyurethane urea in a live canine bladder (5 cm deep) is achieved with 8 seconds of ultrasound-guided HIFU with millimeter resolution energy focus. Tissue sections show no hyperthermic tissue injury. A conceptual application in ureteral stent shape-recovery reduces removal resistance. In conclusion, image-guided HIFU demonstrates deep energy penetration, safety and speed.