Clinical Application and Characteristics of Total Skin Electron Irradiation of Complex Energies

QIAN Jian-sheng,ZHANG Xiao-zhi,TIAN Mei-li
DOI: https://doi.org/10.3969/j.issn.1671-8259.2013.01.029
2013-01-01
Abstract:Objective To compare the characteristics of radiophysics of the muti-fields full body irradiation by 4 Mev and 6 Mev mixed-energy electron beam with the 4 Mev or 6 Mev single-energy electron beam.Methods 3D-water tank,anthropomorphic phantom and dosimeter were used to measure the physical and clinical parameters of 4 Mev or 6 Mev single-energy electron beam or 4 Mev and 6 Mev mixed-energy electron beams generated by Elektra Precise Accelerator,respectively.In addition,the actual measurement results were summarized,analyzed and compared with each other so as to work out the optimal energy allocation of physical and clinical parameters for treating mycosis fungoides.Results When a 5 mm plexiglass scattering screen was applied,the actually measured physical parameters of the 4 Mev and 6 Mev mixed-energy electron beam could meet the clinical requirements.Conclusion The mixed-energy irradiation can not only meet the clinical treatment requirements,but also shorten the treatment time compared to the single-energy electron beam radiation treatment.
What problem does this paper attempt to address?