SPR biosensing method based on phase shift algorithm and its application

Geng Junqing,Yu Xinglong,Zhang Enyao,Dou Fuyin
DOI: https://doi.org/10.3969/j.issn.0254-3087.2012.10.002
2012-01-01
Abstract:A novel surface plasmon resonance (SPR) biosensing method based on phase shift algorithm is introduced in this article. This method comprises a galvo scanning based SPR experimental setup and a phase shift algorithm to determine the resonance angle. A galvo scanner is used to drive the incident laser to carry out angle scanning; thus the interference noise caused by the fan-shape light beam can be avoided. Meanwhile, the galvo scanning is performed step by step, and therefore complete synchronous sampling can be achieved to eliminate the phase noise caused by clock and scanning period jitter. In addition, phase shift algorithm is used to determine the resonance angle, which converts the attenuated total reflection (ATR) curve into frequency domain for processing, the interference of the fluctuation and drift of the light source intensity on measurement result can be reduced. Experimental results show that a refractivity resolution of 1.8×10-6 can be achieved with a linear measurement range of 1.33-1.38, which meets the requirement of biological measurement. Furthermore, SMX/Anti-SMX and MSI-78/SLB interactions were measured using this method. The results demonstrate that this method has a very good application prospect in biomedical study.
What problem does this paper attempt to address?