Equalization of whole-band signal's SNR in the blood components noninvasive measurement

Gang Li,Hui-quan Wang,Hao Zhang,Ling Lin,Xiao-rong Wu,Bao-ju Zhang
DOI: https://doi.org/10.3964/j.issn.1000-0593(2012)02-0486-05
2012-01-01
Spectroscopy and spectral analysis
Abstract:To fully extend the category of blood components that can be noninvasively measured by dynamic spectrum (DS) method and to increase its measuring precision, an overall consideration of light source, tissue absorption and sensor's sensitivity was made. Compensating the light source and adding the telecentric lens not only expand the spectral effective detecting range, but also balance the signal-to-noise ratio (SNR) of the photoelectric pulse in the whole band equalization. The integral SNR of DS signal was increased and the measurement bandwidth was expanded. The effectiveness of this method was validated by the quality evaluation criterion of DS data: the effective detecting range of visible DS was widened from 600 similar to 1 000 nm to 500 similar to 1 135 nm; the effective detecting range of near-infrared DS was widened from 900 similar to 1 100 nm to 900 similar to 1 700 nm. The results show that the design can create the condition for detection of new blood components noninvasively, and enhance the prediction accuracy of the blood components, for which noninvasive measuring using DS method has been achieved.