Quantitative Measurement of Rubidium Isotope Ratio Using Forward Degenerate Four-Wave Mixing

Xuemei Cheng,Zhaoyu Ren,Jian Wang,Yizhu Miao,Xinlong Xu,Lajiang Jia,Haiming Fan,Jintao Bai
DOI: https://doi.org/10.1016/j.sab.2012.04.009
2012-01-01
Abstract:In the present study, forward degenerate four-wave mixing (FDFWM) is employed to measure rubidium (Rb) isotope ratio quantitatively with excellent reproducibility regardless of laser intensity turbulence. The effects of power saturation, atomic absorption and Doppler broadening on measurements are investigated. The results indicate that optimized pump power for isotope ratio measurement is 5-37.3 mW and the optimized atomic number density is around 10(14) m(-3) for the sample cell with a length of 50 mm. In addition, the signal of FDFWM is almost independent on Doppler effect in isotope ratio determination. The measured Rb-85 and Rb-87 ratios in Rb vapor cell and RbCl are 2.590 +/- 0.001 and 2.649 +/- 0.002 respectively, which are in good agreement with the values obtained by absorption spectroscopy and mass spectrometry. A low detection limit as 18.41 fg/mL is achieved. The information obtained from this work suggests that FDFWM is a reliable laser spectroscopic technique in quantitative isotope trace analysis for various applications such as geological and archeological dating. (C) 2012 Elsevier B.V. All rights reserved.
What problem does this paper attempt to address?