Blood compatibility of multi-chamber piezoelectric pump in vitro

Jing SU,Lei ZHOU,Jing-shi DONG,Li-rong GUO,Bai-song LIN
DOI: https://doi.org/10.13481/j.1671-587x.2012.04.015
2012-01-01
Abstract:Objective: To explore the influence of multi-chamber piezoelectric pump on the blood cells and coagulation system when it applied as blood driving pump. Methods: Anticoagulation dog blood was utilized as circulating medium in the simulative circulatory system. There were three groups depended on the circulation time (0, 1 and 2 h) and a control group. Hemoglobin concentration, white blood cell, red blood cell and platelet counts, hematocrit, free hemoglobin concentration were detected. The index of hemolysis (IH) of multi-chamber piezoelectric pump was calculated as an important index of hemolysis damage. Results: Compared with control group, each index was not significantly changed in cycle 0 h group(P>0.05). Compared with control group, the red blood cell count, white blood cell count, platelet count, hemoglobin concentration, hematocrit were decreased in cycle 1 h and 2 h groups (P<0.05). Compared with control group, the free hemoglobin concentration was increased, the fibrinogen concentration was decreased(P<0.05). Compared with cycle 1 h group, each index was not significantly changed in cycle 2 h group(P>0.05). The IH of pump was calculated as 0.004 5, and it was significantly lower than international standard (<0.1). Conclusion: Multi-chamber piezoelectric pump has good blood compatibility in vitro. It has advantages to apply as blood-driving pump.