Quantitative Determination of Cyclophosphamide in Rat Plasma Using an On-Line SPE HPLC-DAD

Xiao-Na Li,Ning Yu,Jian-Mei Zhang,Wen-Si Lin,Xiao-Mei Ling,Ge Fu,Run-Tao Li,Jing-Rong Cui
DOI: https://doi.org/10.5246/jcps.2012.02.020
2012-01-01
Journal of Chinese Pharmaceutical Sciences
Abstract:A rapid and simple liquid chromatography method with on-line solid phase extraction was developed and validated for the quantitative determination of cyclophosphamide in rat plasma.The plasma sample was first extracted on an Acclaim(R) Polar Advantage Ⅱ C18 guard column (PA Ⅱ C18,10 mm×4.6 mm,5 μm),which was also the on-line Extraction Cartridge SPE column,by washing with 100% H2O for 1 min.The extracted sample was then eluted onto a PA Ⅱ C18 column (150 mm×4.6 mm,5 μm) and separated by isocratic elution with acetonitrile-water (40∶60,v/v).The mobile phase was run at a flow rate of 1.0 mL/min,and the UV detector was set at 195 nm.Retention time of cyclophosphamide was 4.3 min and the total run-time was 6 min.The linear range of the standard curve was from 1.0 to 200 μg/mL (r2 =0.9999),and the limits of quantification and detection were 1.0 μg/mL (RSD<10%,n =5) and 0.3 μg/mL (RSD<13%,n =5),respectively.Both intra- and inter-day variations were less than 5.6%.The developed method can be used for the therapeutic drug monitoring of cyclophosphamide in the clinic.
What problem does this paper attempt to address?