Characteristic Investigation of a Static Micro Polymerase Chain Reaction Chip Based on in Situ Electrochemical Detection

Tao Deng,Mengwei Li,Chenxu Zhao,Jian Qin,Cangran Guo,Zewen Liu
DOI: https://doi.org/10.1049/mnl.2012.0551
2012-01-01
Micro & Nano Letters
Abstract:Aiming to meet the ever-increasing challenge of a low cost, portable and real-time sensitive detection polymerase chain reaction (PCR) system, a novel in situ electrochemical detection static micro PCR (EC-S mu PCR) chip is designed and fabricated. The 10 x 10 mm EC-S mu PCR chip comprises 12 micro reaction chambers, and each 1500 x 1000 x 60 mu m chamber consists of three modules: a micro heater, a micro temperature sensor and a micro three-electrode system (MTES). Heating, cooling and electrochemical cyclic voltammetry (CV) experiments were carried out to investigate the performance characteristics of each module of the chamber in the EC-S mu PCR chip. The results showed that the heating rate of the chip was 1.05 degrees Cs-1 with only one micro heater, whereas the cooling rate was 1.01 degrees Cs-1 with natural air cooling. It is worth noting that the bare MTES of the chamber directly detected the oxidation of single-stranded DNA (ssDNA, 10(-6) M) with the CV method, furthermore, the oxidation peak current of ssDNA can be increased about 40% after a prepolarisation step. All three modules of the chamber in the EC-S mu PCR chip work efficiently and reliably, which reveals that the entire chip is ready for further analysis trials.
What problem does this paper attempt to address?