Numerical analysis of electrical impedance scanning imaging of breast

Zhang Feng,Luo Limin,Bao Xudong,Chen Beijing,Zhang Yue
DOI: https://doi.org/10.3969/j.issn.1001-0505.2012.05.014
2012-01-01
Abstract:The existing models for analyzing electrical impedance scanning imaging (EISI) may produce results significantly different from real clinic findings. According to women's breast structure, Chinese women's breast size, electric parameters of breast, and the structure of EISI probe, an improved EISI model is presented. Statistic analysis of EISI parameters is presented by the finite element method (solving quasi-static equation of the imaging model) and the Yates method (analysis of variance). Results indicate that the measured current at breast surface can display corresponding disturbance when there is cancerous lesion within breast due to different conductance in breast cancerous tissue and surrounding normal tissue. Degree of current disturbance is related with the breast size, the depth and radius of cancerous lesion, the conductance of surrounding tissue of cancerous lesion, and the interaction between above parameters. If the breast size and the depth of cancerous lesion can be decreased by measurement jig, the detection performance of breast cancer can be improved.