Electrical Impedance Spectroscopy for Ex-Vivo Breast Cancer Tissues Analysis

Francesco Meani,Giuseppe Barbalace,Davide Meroni,Olivia Pagani,Ulrike Perriard,Alberto Pagnamenta,Andrea Aliverti,Emanuele Meroni
DOI: https://doi.org/10.1007/s10439-023-03159-4
Abstract:This paper describes a cancer detection procedure based on customised electrical impedance spectroscopy (EIS) in breast cancer surgical samples and an analysis of its outcomes. A tissue analyser was developed to inject a low-amplitude alternating current with penetrating electrodes into breast specimens along a broad spectrum of frequencies. Experimental measurements were carried out on more than one hundred excised breast cancer specimens, with the goal of discriminating between the tumour and surrounding non-neoplastic tissue. The probe was inserted in different locations immediately after surgical excision in order to measure tissue impedance (modulus and phase). Electrical impedance varied significantly between neoplastic and surrounding non-neoplastic tissues, with a low standard deviation among the different measurements, confirming good reproducibility. Tumours could be discriminated from non-neoplastic tissues according to their impedance modulus value for high frequencies and phase value for low frequencies. Impedance also varied significantly in both non-neoplastic and tumour tissues depending on the patient's age and tumour characteristics, such as size and histological sub-type. EIS is able to discriminate between healthy tissue and cancer. Future developments of this technology could be exploited for intraoperative real-time evaluation of the transition zone between cancer and normal tissues.
What problem does this paper attempt to address?