Design of Automatic Analysis Instrument for Cell Physiological Multi-Parameter Detection Based on Integrated Chip

WU Cheng-xiong,CAI Hua,HU Ning,HU Zhao-ying,CHENG Gong,XIAO Li-dan,YU Hui,WANG Ping
DOI: https://doi.org/10.3785/j.issn.1008-973x.2012.09.026
2012-01-01
Abstract:An automatic analysis instrument was developed to simultaneously record the multi-physiological parameters of the cells.The instrument employs an integrated chip which consists of light addressable potentiometric sensor(LAPS),electrical cell-substrate impedance sensor(ECIS) and microelectrode array sensor(MEAS),three different types of cell-based biosensors,and their detection systems.A cell growth environment control system and an automatic flow analysis system were integrated in the instrument,so it can meet different needs of cell physiology experiments.Compared with previous cell physiology testing instruments,the new instrument comes to be multi-parameter,multi-function,automatic and flexible.An investigation of extracellular acidification rate change and doxorubicin inducing change of cell impedance and extracellular potential was carried out.Result shows that the instrument is capable of monitoring extracellular ions,cell impedance and extracellular potential in synchronous and long-time.Thus it can be used in drug evaluation,drug screening and other areas of biomedical research.