Implantable Pt-Black Coated Microelectrode for Electrical Stimulation

RUI Yue-feng,LIU Jing-quan,YANG Chun-sheng
DOI: https://doi.org/10.3969/j.issn.1672-6030.2012.02.002
2012-01-01
Abstract:Conventional implantable electrical stimulation microelectrodes have small surface area,high electrode-tissue interface impedance and low charge storage capacity(CSC),which will increase power consumption and hinder stimulation effect.In this paper,a method was presented to increase the effective surface area(ESA) of microelectrode by coating Pt-black on the electrode sites.The method combines with electroplating Pt-black by pulse current under ultrasonic bath,which can significantly increase the ESA,decrease the interface impedance,and increase the CSC and charge-injection capacity.Moreover,the geometrical properties of Pt-black coated microelectrodes were demonstrated by using a scanning electron microscope(SEM) and the electrical properties were measured by an electrochemical analyzer compared with conventional microelectrodes without coatings.The mechanical stability of Pt-black coating was also tested.Experimental results show that approximately 16 times lower interface impedance and 13 times higher CSC were achieved by this nanostructured microelectrode.Cathodal charge storage capacity(CSCC) decreased only by 20% after an ultrasonic bath at room temperature for 5 min.
What problem does this paper attempt to address?