Determination of 3-Phenoxybenzoic Acid in Microbial Degradation Systems by HPLC-UV Detection

ZHAO Nan,LIU Shu-liang,LAI Wen,YUAN Huai-yu
2011-01-01
Abstract:A HPLC method with UV detection was proposed to determine the concentration of 3-phenoxybenzoic acid(3-PBA) in microbial degradation systems.The chromatographic separation was performed on a Gemini 100A C18 column(5.0μm,150 mm × 4.60 mm,i.d.) using a mobile phase made up of acetonitrile and water(70:30,V/V) at a flow rate of 0.7 mL/min,and 3-PBA was detected at 210 nm using a UV detector.The retention time of 3-PBA was 4.268 min.The linear range of the method was 0.5-50.0 mg/L.The average spike recovery for 3-PBA was 98.919%,with a relative standard deviation(RSD) of 2.78%.The residues of 3-PBA in 4 different microbial degradation systems were 24.467,86.266,2.633 mg/L and 1.921 mg/L,respectively.The developed method demonstrated the benefits of simplicity,rapidity,high accuracy and good separation.