A $160~\mu {\rm W}$ 8-Channel Active Electrode System for EEG Monitoring

Jiawei Xu,Refet Firat Yazicioglu,Bernard Grundlehner,Pieter Harpe,Kofi A. A. Makinwa,Chris Van Hoof
DOI: https://doi.org/10.1109/tbcas.2011.2170985
2011-01-01
IEEE Transactions on Biomedical Circuits and Systems
Abstract:This paper presents an active electrode system for gel-free biopotential EEG signal acquisition. The system consists of front-end chopper amplifiers and a back-end common-mode feedback (CMFB) circuit. The front-end AC-coupled chopper amplifier employs input impedance boosting and digitally-assisted offset trimming. The former increases the input impedance of the active electrode to 2 GΩ at 1 Hz and the latter limits the chopping induced output ripple and residual offset to 2 mV and 20 mV, respectively. Thanks to chopper stabilization, the active electrode achieves 0.8 μVrms (0.5-100 Hz) input referred noise. The use of a back-end CMFB circuit further improves the CMRR of the active electrode readout to 82 dB at 50 Hz. Both front-end and back-end circuits are implemented in a 0.18 μm CMOS process and the total current consumption of an 8-channel readout system is 88 μA from 1.8 V supply. EEG measurements using the proposed active electrode system demonstrate its benefits compared to passive electrode systems, namely reduced sensitivity to cable motion artifacts and mains interference.
What problem does this paper attempt to address?