Label-free Detection of Clenbuterol Based on Microcantilever

HUANG Yuan,XUE Chang-guo,ZHANG Qing-chuan,TAN Wei-ming,WANG Bao-min,WU Xiao-ping
2009-01-01
Abstract:Objective A novel technology of microcantilever sensor platform for label-free detection of clenbuterol antigen-antibody specificity binding is reported. Method Thiolated clenbuterol antibody was immobilized on the gold-coated side of the microcantilever by self-assembled monolayer technique.Consequently the deflections of the microcantilever corresponding to the process of injecting different concentration of clenbuterol antigen were real-time monitored by optical lever technique.The activity of thiolated clenbuterol antibody and the clenbuterol antigen-antibody binding on the gold-coated side of the microcantilever were confirmed by an Enzyme-Linked Immunosorbent Assay(ELISA). Result The sensitivity of detecting clenbuterol is better than 1 μg/L.Microcantilevers functionalized with thiolated clenbuterol antibodies were exposed to 1 mg/L of Chloramphenicol(CAP),and no significant response was detected,which confirmed the specificity of the immobilization method.In addition,clenbuterol antigen-antibody binding on the microcantilever generates a compressive surface stress and the surface stress is linear to logarithm of clenbuterol concentration over a range of concentration from 1ng/mL to 1 μg/mL. Conclusions It is possible for detection of clenbuterol by microcantilever sensor platform.
What problem does this paper attempt to address?