Resting Pd/Pa measured with intracoronary pressure wire can predict fractional flow reserve

CUI Ming,YAN Li-qiu,ZHANG Yong-zhen,WANG Xin,ZHANG Fu-chun,NIU Jie,LI Hai-yan,WANG Gui-song,HAN Jiang-li,GAO Wei,GUO Li-jun
DOI: https://doi.org/10.3969/j.issn.1004-8812.2011.01.003
2011-01-01
Abstract:Objective To investigate the relationship between ratio(Pd/Pa)of resting distal coronary pressure to aortic pressure and fractional flow reserve(FFR)obtained during maximal hyperemia.Methods One hundren and three lesions with pressure-wire studies performed in 84 patients were retrospectively analyzed.Results A linear correlation between resting Pd/Pa and FFR post-pharmacological hyperemia was observed(r=0.78;P<0.0001).When a FFR of ≤0.75 was defined as a cut-off value of ischemia,a resting Pd/Pa of ≤0.86 had a positive predictive value(PPV)of 95%,while a resting Pd/Pa of ≥0.93 had a negative predictive value(NPV)of 95%.Conclusions We demonstrate a strong correlation between resting Pd/Pa and FFR.Resting values of Pd/Pa can be used to predict the cut-off value of ischemia of FFR with relatively high PPV and NPV.This may potentially obviate the need for adenosine or other hyperemia agents infusion in a proportion of pressure-wire studies,reduce adverse reactions of drugs and shorten the process of operation.
What problem does this paper attempt to address?