Rapid Detection of Tetramethylenedisulfotetramine in Human Blood by Solid-Phase Microextraction/gas Chromatography

TG Luan,GK Li,MQ Zhao,ZX Zhang
DOI: https://doi.org/10.1016/s0003-2670(99)00725-4
IF: 6.911
2000-01-01
Analytica Chimica Acta
Abstract:The solid-phase microextraction (SPME) coupled with gas chromatography (GC) was used to develop a simple, rapid and sensitive method for the determination of tetramethylenedisulfotetramine (tetramine) in human blood. The tetramine was extracted with a fused silica fibre coated with a 100μm polydimethylsiloxane (PDMS) and detected by GC with a nitrogen/phosphorus detector (NPD). The effects of extraction and desorption times as well as the extraction temperature were studied. A linear response of tetramine over a concentration range of 0.008–0.5μgml−1 with a correlation coefficient of 0.998 and detection limit 0.001μgml−1 (3s criteria) were obtained. Spiked human blood samples were analyzed. The analytical recoveries ranged from 92 to 114% and the mean RSD was 10.5%.
What problem does this paper attempt to address?