Quartz Crystal Biosensor for Real-Time Kinetic Analysis of Interaction Between Human TNF-α and Monoclonal Antibodies

Y Liu,W Zhang,X Yu,HW Zhang,R Zhao,D Shangguan,Y Li,BF Shen,GQ Liu
DOI: https://doi.org/10.1016/j.snb.2003.12.010
IF: 9.221
2004-01-01
Sensors and Actuators B Chemical
Abstract:Quartz crystal microbalance (QCM) biosensor integrated in a flow injection analysis (FIA) system was used for the real-time investigation of the interaction between human TNF-α (hTNF-α) and its monoclonal antibodies (mAbs). Biologically active hTNF-α trimer was immobilized on the gold electrodes of the piezoelectric crystals and the binding interactions of the two immobilized antigen with two anti-hTNF-α monoclonal antibodies (Z8 and Z12) were followed as changes in the resonant frequency of the modified crystals. The resultant sensorgrams were rapidly analyzed by using an in-house kinetic analysis software based on genetic algorithm (GA) to derive both the kinetic rate constants (kass and kdiss) and equilibrium dissociation constant (KD) for antigen–antibody interactions. For the interactions, KD were 2.31×10−7 and 5.42×10−7 M for Z8 and Z12, respectively. The generally good agreement found between biosensor data and the affinities known from conventional equilibrium enzyme-linked immunosorbent assay (ELISA) validates this analytical method. The study shows the potential usefulness of QCM biosensor technology for biomolecular interaction analysis.
What problem does this paper attempt to address?