Determination of SOD by Electrochemiluminescence Quenching Analysis

CHU Hai-hong,WU Ying,DI Jun-wei,TU Yi-feng
DOI: https://doi.org/10.3969/j.issn.1004-4957.2006.01.034
2006-01-01
Abstract:It was found that the sensitized luminol electrochemiluminescence(ECL) formed through the reaction of dissolved oxygen or hydrogen peroxide with luminol at a boric acid buffer solution of pH 6.7,could be quenched by SOD.A sensitive and reliable method based on the above phenomenon was thus established for the determination of SOD.The sensitized ECL intensity has a linear relationship with the SOD concentration in the range of 0.24-9.6 IU/mL for luminol-O_2 system,and in the range of 0.024-1.8 IU/mL for luminol-H_2O_2 system.The mechanism of the quenching effect of SOD on the sensitized luminol ECL was also studied.