Abstract:The residual silanols activity of the stationary phase surface could not be characterized by conventional methods for evaluating reversed-phase High Performance Liquid Chromatography(HPLC) columns.Therefore,a new method for characterizing reversed-phase HPLC columns in terms of the silanols activity using peak asymmetry factor of amitriptyline as an appropriate measure has been developed in this paper.Ten different types of reversed-phase columns were evaluated by this new method.According to the evaluating results,two types of columns were selected to separate and analyse basic antibacterial agents.The results showed that this new method could effectively characterize silanols activity of reversed-phase HPLC packing and some significant information could be offered by the evaluation results for the selection of columns to analyse organic bases.