Glucose biosensor based on platinum nanowire array

Wang Cun-chang,Lu Ya-shuang,Yang Ming-hui,Wu Guo-qiang,Shen Guo-li,Yu Ru-qin
DOI: https://doi.org/10.3969/j.issn.1008-2298.2007.03.005
2007-01-01
Chemical Sensors
Abstract:Platinum nanowire arrays can be grown by electrodeposition in polycarbonate membrane,with the average diameter of the nanowires about 250 nm and the length about 2 μm.SEM characterization indicates the nanowire array has an ordered and uniform structure with nanowire density of 5×108 cm2.The electrochemical behavior of the nanowire modified electrode was investigated.The platinum nanowire array modified electrode can direct response to hydrogen peroxide at low potential of-0.1 V with wide linear range(1×10-7 to 5×10-2 mol/L)and high sensitivity,indicating excellent electrocatalytic activity of the modified electrode.With the immobilization of glucose oxidase onto the electrode surface using glutaric dialdehyde,a new glucose biosensor has been constructed.Interference free determination of glucose has been realized with a linear range from 5 μmol/L to 20 mmol/L and the detection limit was 1 μmol/L glucose(S/N=3).
What problem does this paper attempt to address?