Calibration Apparatus for Artillery Chamber Pressure Test Device Based on Environmental Factors

ZHANG Yu,ZU Jing,ZHANG Hongyan,MA Tiehua
2012-01-01
Abstract:In order to overcome the disadvantage of the simulation chamber calibration apparatus,such as complex structure,inconvenience and pressure range is too small,we improved the existing apparatus and established a new calibration system.The calibration system consisted of simulation chamber calibration apparatus,standard testing system,high and low temperature test equipment.The artillery chamber pressure testing device could be calibrated by adopting environmental factors.The improved calibration apparatus could simulate the environment of high temperature and high pressure in gun chamber,the amplitude of the pressure signal generated by the apparatus was between 80~800 MPa,the pulse width of the pressure signal was between 10~20 ms,the rising time was less than 2 ms.Theory analysis and contrast test results showed that the calibration method for simulating application environment of it was reasonable.The sensitivity coefficients obtained by static calibration and calibration of simulation application environment were different.The practical firing results showed that the data from the testing device after simulation environment calibration had better stability and high confidence level.