Assessment of the Accuracy of Continuous Glucose Monitoring and Its Correlated Factors

LU Wei,ZHOU Jian,MA Xiao-jing,PAN Jie-min,ZHU Wei,LU Feng-di,BAO Yu-qian,JIA Wei-ping
2011-01-01
Abstract:Objective To assess the accuracy of continuous glucose monitoring(CGM) for patients with type 2 diabetes mellitus(T2DM),and its correlated factors.Methods From January 2009 to January 2010,530 patients with T2DM were enrolled in this study,including 265 subjects taking oral hypoglycemic agents and the other 265 subjects taking intensive insulin treatment.All the subjects underwent CGM(American Medtronic MiniMed) for three days.Meanwhile,capillary glucose values through self-monitoring of blood glucose(SMBG) were inputted four times a day for adjustment.The correlation of CGM value and capillary glucose value was analyzed by linear regression method.The consistency was analyzed by Clarke error grid.Mean absolute relative difference(MAD) was used to assess accuracy and correlated factors of MAD were also analyzed.Results ① A total of 6 350 pairs of CGM and SMBG values were collected [(9.66±3.54) mmol/L vs.(9.64±3.38) mmol/L,P>0.05].Pearson correlation analysis showed that CGM value was positively correlated with SMBG value(r=0.959,P<0.001).② Clarke error grid demonstrated that 99.89% of paired SMBG-CGM values were located in zone A and zone B and the remaining seven pairs(0.11%) of glucose values were located in zone D.③ The overall MAD value was 7.2%(5.5%-9.5%).According to the analysis of daily MAD value,MAD value decreased,while accuracy elevated gradually with monitoring time.④ MAD value of intensive insulin treatment group was higher than that of the oral hypoglycemic agent treatment group(P<0.05).And stepwise multiple regression analysis indicated that MAD value had negative correlation with glycated albumin level(β=-0.134,P<0.01) and positive correlation with intensive insulin treatment(β= 0.117,P<0.05).Conclusions ① The results of CGM are accurate,and have good correlation and consistency with capillary glucose.② Besides monitoring time,the accuracy of continuous glucose monitoring may be also associated with hypoglycemic treatment strategy.
What problem does this paper attempt to address?