Accuracy of a Continuous Glucose Monitor During Pediatric Type 1 Diabetes Inpatient Admissions

Erin Cobry,Laura Pyle,Lauren Waterman,Gregory P. Forlenza,Lindsey Towers,Angela Karami,Emily Jost,Cari Berget,R. Paul Wadwa
DOI: https://doi.org/10.1089/dia.2023.0375
2024-01-11
Diabetes Technology & Therapeutics
Abstract:Objective Continuous glucose monitors (CGM) used for type 1 diabetes management are associated with lower HbA1c. CGMs are not approved for inpatient use, when close glucose monitoring and intensive insulin management are essential for optimal health. Accuracy data from adult hospitalizations have been published but pediatric data is limited. Design and Methods This retrospective review of Dexcom G6 data from youth with type 1 diabetes during hospitalization assessed CGM and matched (within 5 minutes) point of care (POC) and lab glucose values. Glucose values >400 mg/dl and 100mg/dl and 15/20/30mg/dl if value ≤100mg/dl. Results A total of 1,120 POC and 288 lab matched pairs were collected from 83 unique patients (median age 12.0 yrs, 68.7% non-Hispanic white, 54.2% male) during 100 admissions. For POC values, overall MARD was 11.8%, on the medical floor was 13.5% and in the ICU was 7.9%. The MARD for all lab values was 6.5%. 98% of matched pairs were within Clarke Error Grid A and B zones. Conclusions Findings from our pediatric population were similar to accuracy reported in hospitalized adults, indicating the potential role for CGM use during pediatric hospitalizations. Additional research is needed to assess accuracy under various conditions, including medication use, as well as development of safe hospital protocols for successful CGM implementation for routine inpatient care.
endocrinology & metabolism
What problem does this paper attempt to address?