Determination of Chenodexycholic Acid Using Pre-column Derivatization and Reversed-phase High Performance Liquid

郑兴,金东日
DOI: https://doi.org/10.3969/j.issn.1004-4353.2008.04.012
2008-01-01
Abstract:Using p-bromoacetyl-nitrobenzene as pre-column derivatization reagent,the method was studied to determine the content of chenodeoxycholic acid by reversed-phase high performance liquid Chromatography.The result showed that the best condition for the reaction was at 55℃ and 60 minutes.Chromatography analysis,using methanol and water as mobile phase,was carried out at 265nm.The linearity range was within 0.005 1~0.306 0 g/L.The precission was good with RSD less than 1.2% and the limit of detection was 4.0ng.The reaction was fast and the derivative was stable;the analysis was accurate and sensitive.It was suitable for determination of chenodeoxycholic acid.