Abstract:Objective: To establish a high performance liquid chromatography method using pre-column derivation for the test of protein residue in chitosan.Methods:Chitosan samples were hydrolyzed by hydrochloric acid and derived using 9-fluorenylmethylchloroformate(FMOC-Cl).The analysis was carried out on a C18 column using 5 mmol·L-1 ammonium acetate buffer(pH 6.7)-acetonitrile(80∶ 20)and 5 mmol·L-1 ammonium acetate buffer(pH 6.7)– acetonitrile(20∶ 80)as mobile phases with linear gradient elution and a fluorescence detection(excitation 266 nm,emission 313 nm).Amino acids were calculated using external standard method and the total amino acid was counted as the protein residue in chitosan.Results:Good retention and separations were achieved for derivatives of 16 amino acids.Only Leu,Ile and Phe were interfered by the excess derivation reagents and derivation products of glucosamine.The calibrations of amino acids were linear(r>0.99)within 0.01-5.0 μg·mL-1,and the precision(RSD),recovery and limit of quantification(LOQ)of the established method were 1.6%-8.5%,84.4%-98.6%,and 1.9-10.9 ng·mL-1,respectively.Conclusion:The established method can accurately quantify 13 amino acids in chitosan and thus can be used for the test of protein residue in chitosan for quality evaluation of chitosan.The method could provide reference for guarantee of the safety for the use of chitosan in medical and biological materials.