Clinical investigation of sensitivity and specificity of dust mite skin prick test kit

GU Hongliang,QIAN Yaoqin,WANG Peihua,CAO Lanfang,XUE Haiyan,ZHENG Hongliang
2008-01-01
Abstract:Objective To evaluate the sensitivity and specificity of dust mite skin prick test kit in diagnosis of dust mite allergic asthma and rhinitis.Methods A multi-centered parallel trail was performed in 150 dust mite allergic patients and 150 non-dust mite allergic patients using dust mite skin prick test kit.The hypersusceptibility degrees of patients to dust mite were evaluated by comparing the areas caused by mite prick test and positive control solution.The sensitivity and specificity of the skin prick test kit was the major task of evaluation.Results The sensitivity and specificity of the kit for dust mite allergy were 93.3%and 96.0%,respectively.The area under receiver operating characteristic (ROC)curve was 0.950 8.The patients showed a good compliance to the kit and there was no severe adverse events happened in the trial.Conclusion The dust mite skin prick test kit is safe,highly sensitive,highly specific and accurate;it can be used as assistant diagnosis method for allergic disease and provides reliable evidence for targeted immunotherapy.