The Calibration Model Optimization of Gross Energy in Feed by Near Infrared Reflectance Spectroscopy(NIRS)

Zhiyou Niu
DOI: https://doi.org/10.3969/j.issn.1003-188x.2008.08.007
2008-01-01
Abstract:The use of near Infrared reflectance spectroscopy(NIRS) was investigated as an alternative method for predicting gross energy in feed. The intial spectrum was pretreated by multiplicative scantter correction(MSC),standard norma lized variate(SNV),first derivative and second derivative ,respectively.The NIRS calibration models were developed using the partial least squares(PLS) technique. The methods of standard normalized variate(SNV) and first derivative were chosen as the best pretreatment. The coefficient of determination in calibration (R2 ) and standard error(RMSEC) of chemical analysis value and NIRS prediction value of calibration set were 0.9060 and 0.153.The coefficient of determination in calibration (r2 ) and standard error(RMSEP) of validation set were 0.892 4 and 0.156,and the RPD were all over 3. It was concluded that NIRS can be used as a method to monitor the chemocal composition and gross energy of feed for ruminant, and a good measuring precision could be expected.
What problem does this paper attempt to address?