Abstract:Objective To explore the value of color Doppler flow imaging (CDFI) and contrast-enhanced ultrasound (CEUS) in detecting hepatic artery complications in patients after liver transplantation and to compare the usefulness of CDFI and CEUS.Methods A total of 267 CDFI examinations for 158 patients after liver transplantation were studied.Data were collected with attention to the patency,peak velocity,acceleration time and resistance index of intra- and extrahepatic artery,as well as turbulence of extrahepatic artery.Thirty-three patients suspected by CDFI underwent CEUS with SonoVue and Contrast Tuned Imaging (CnTI) to observe hepatic artery and perfusion of liver parenchyma.Results In 19 patients,17 patients were with hepatic artery stenosis (HAS) and two with hepatic artery occlusions (HAO) confirmed by DSA and/or CT angiography (CTA).Vascular parameters of 17 patients with HAS were abnormal with sensitivity of 100%,but CDFI could not show the stenosis position and degree.CDFI did not demonstrate flow signal in the 2 patients with HAO,and could not make a definite diagnosis.CEUS confirmed not only HAO,but also displayed the stenosis position and degree.The accordance rates for HAO,HAS detected by CEUS compared to DSA and/or CTA were 100% (2/2),94.1% (16/17) and 88.2% (15/17),respectively.Conclusions CDFI complements CEUS mutually,which may improve the diagnostic accuracy of hepatic artery complications after liver transplantation.