Comparative Study on Three Testing Methods for Urine Formed Elements
ZHU Cheng-bin,QU Chao,DOU Lu,GAO Ying-dong,HE Bang-shun
DOI: https://doi.org/10.13241/j.cnki.pmb.2010.19.027
2010-01-01
Abstract:Objective: To investigate the coincidence rate and influence factors by using three different methods of examination of Urine formed elements. Method: Examine the urine by AX-4280 automatic dry chemistry analyzer, IQ200 automated urine microscopy analyzer and method of artificial joint detection and compare and analyze the results. Results: Urine test results of various physical components exist in different degrees of false positives, artificial microscopic examination of compliance with the AX-4280 was: 71.1% RBC; WBC 76.4%. Manual examination method and IQ200 automated urine analyzer manual modification before and after the artificial modification of the results found were: RBC 81.3%, 91.5%; WBC 85.9%, 93.8%; transparent tube 81.4%, 89.7%, uncategorized tube 52.6%, 77.4%; crystallization 84.1%, 92.4%; bacteria 65.1%, 88.2%; sperm 72.0%, 91.5%. Cystoscopy was the standard artificial, IQ200 automated urine analyzer manual modification before and after the artificial modification of test items of various false positive rates were: 23.0% RBC, 9.3%, 16.4% WBC, 6.6%, 22.9% transparent tube, 11.4%, 90.2% uncategorized tube, 29.2%, crystalline 19.0%, 8.2% and 53.7% of bacteria, 11.0%, 38.9% of sperm, 9.3%. Conclusion: The combined application of three methods of urine formed elements can enhance the accuracy of test results, provide a basis for clinical diagnosis and treatment.