Design and Fabrication of PIN-PMN-PT Single-Crystal High-Frequency Ultrasound Transducers.

Ping Sun,Qifa Zhou,Benpeng Zhu,Dawei Wu,Changliong Hu,Jonathan M. Cannata,Jin Tian,Pengdi Han,Gaofeng Wang,K. Kirk Shung
DOI: https://doi.org/10.1109/tuffc.2009.1367
2009-01-01
Abstract:High-frequency PIN-PMN-PT single crystal ultrasound transducers at center frequencies of 35 MHz and 60 MHz were successfully fabricated using lead indium niobate-lead magnesium niobate-lead titanate (0.23PIN- 0.5PMN-0.27PT) single crystal. The new PIN-PMN-PT single crystal has higher coercivity (6.0 kV/cm) and higher Curie temperature (160 degrees C) than PMN-PT crystal. Experimental results showed that the PIN-PMN-PT transducers have similar performance but better thermal stability compared with the PMN-PT transducers.
What problem does this paper attempt to address?