Standard particle experimental validation of cutpoint diameters on novel cascade impactor

Liu Yunxi,Chen Xu,Yang Fumo
DOI: https://doi.org/10.3969/j.issn.1673-9108.2006.05.032
2006-01-01
Abstract:Aerosols affect greatly the environment pollution,climate changes and human health.The study of aerosols sampling and analyzing becomes an important object in the atmospheric environment research fields.Novel cascade impactor(NCI) provides an efficient instrument for sampling aerosol particles separated into 7 stages according to their sizes.The nearly even and flat samples obtained with different particles are propitious to the physical and chemical analysis subsequently.Its assurance for efficiency and stabilization of cutpoint diameters on every stage has been tested by experiments.Their cutpoint diameters and overall efficiency on the first three stages have been experimentally validated by the uniform and mono-disperse standard particles with same surface characteristics,provided by the vibrating orifice aerosol generator.The results indicate that NCI provides different particles collection for sampling aerosols according to their sizes.