Linearity research of Cu&Pb in tea by inductively coupled plasma-atomic emission spectroscopy

ZHENG Wei-dong,YANG De-gui,ZHENG Shi-chao,HU Dan,ZHOU Ya,CHENG Gui-hong,LIN Hong
DOI: https://doi.org/10.3969/j.issn.1004-1656.2013.10.021
2013-01-01
Abstract:To provide systematic reference for the linear performance rescerch of detection methods in the spectral analysis field,according to Yunnan province standard DB53/T 288-2009,Cu&Pb in the tea were detected by LCP-AES,studied standard solution,reagent blank and sample digestion fluid effect of three substrates on the calibration line properties. During the method validation stage,implementation calibration line performance study,to find the fittest calibration linear range by study the higher&lower pot not follow Lambert-Beer and taking the content of target component in the sample into account. The first low concentration pot after zero pot should far away the limit of detection of the method,and to ensure accurate and reliable testing results,it's important to keep the component concentration of the sample digestion fluid in the best working linear range of calibration linear.