Impact of the Number of Image Planes of Real-Time Three-Dimensional Echocardiography on the Accuracy of Left Atrial and Ventricular Volume Measurements

Fang Li,Qian Wang,Gui Hua Yao,Peng Fei Zhang,Zhi Ming Ge,Mei Zhang,Yun Zhang
DOI: https://doi.org/10.1016/j.ultrasmedbio.2007.07.012
IF: 3.694
2008-01-01
Ultrasound in Medicine & Biology
Abstract:Real-time three-dimensional (3D) echocardiography (RT-3DE) has emerged as a new technique in measuring left atrial and ventricular volume. However, the impact of cutting planes of RT-3DE on the accuracy of volume measurement in patients with a normal or enlarged heart is still unknown. We enrolled 30 normal subjects (control group) and 30 patients with heart failure (patient group). RT-3DE was performed to measure maximal volume of the left atrium (LAVmax) and left ventricular end-diastole volume (LVEDV) with 2-, 4-, 8- and 16-cutting planes, compared with cardiac magnetic resonance imaging (CMRI). In both groups, LAVmax by RT-3DE using 2- and 4-cutting planes was significantly underestimated (mean difference: -10.4 +/- 16.6 mL, p = 0.001 and -8.8 +/- 14.2 mL, p = 0.002 in the control group and -13.4 +/- 19.6 mL, p = 0.001 and -11.2 +/- 17.5 mL, p = 0.001 in the patient group, respectively). These differences became nonsignificant when 8- and 16-cutting planes were adopted (mean difference: -2.1 +/- 7.6 mL and -1.9 +/- 7.4 mL in the control group and -2.7 +/- 8.4 mL and -2.2 +/- 8.3 mL in the patient group, respectively). The agreement for LVEDV was acceptable when 4- or more cutting planes were used in the control group and when 8- or 16-cutting planes were used in the patient group. The time expense for data analysis of LAVmax with 8-image planes was only 7 +/- 4 min in the control group and 6 +/- 5 min in the patient group, almost halving that of the 16-image planes. Similarly, 4- and 8-cutting planes were required for an accurate measurement of LVEDV in the control and patient groups, respectively. In conclusion, RT-3DE with 8-cutting planes is both accurate and timesaving for measurement of LAVmax and LVEDV in patients with normal or enlarged left atria and ventricles.
What problem does this paper attempt to address?