Comparison of the Effect of LAMP and RT-PCR Assays for Rapid Detection of Mycobacterium Tuberculosis in Sputum Samples

BAO Wei-hua,LIU Yi,HUANG Shu-hai,LAN Lan,ZHOU Chang-ming,LIAO Rui-qing,YAO Yi-qiang
DOI: https://doi.org/10.3969/j.issn.0253-4304.2012.02.010
2012-01-01
Abstract:Objective To compare the effect of loop-mediated isothermal amplification(LAMP)and real-time reverse transcription polymerase chain reaction(RT-PCR)assay for the detection of mycobacterium tuberculosis in sputum samples.Methods One hundred and eighteen sputum samples from suspicious tuberculosis(TB)patients were detected by the four methods including smear microscopy,L-J solid culture,LAMP and RT-PCR assay.Results The positive detection rates of smear microscopy,L-J culture,LAMP and RT-PCR were 38%,41%,45% and 55%,respectively.There were significant differences in the positive rate among the four methods(P0.05).Compared to the L-J solid culture,the sensitivity and specificity of LAMP were 92.0%(46/50)and 86.8%(59/68),and there was no significant difference between the two methods(P0.05),but with high concordance(Kappa=0.777).The sensitivity and specificity of RT-PCR assay were 86.0%(43/50) and 66.2%(45/68),and showed significant differences between the L-J solid culture and RT-PCR assay(P0.01).Conclusion RT-PCR assay shows the highest positive rate in the four methods,but of lowest specificity.LAMP may provide a simple,rapid,sensitive and cost-effective method for the detection of pulmonary TB,with high sensitivity and specificity,which has a high concordance with traditional culture.
What problem does this paper attempt to address?