Quantitative study of ultrasonic vector flow mapping in experimental hydrody-namic model

李小庆,唐红,李晨,周璨,李海华
DOI: https://doi.org/10.3969/j.issn.1672-6448.2009.05.004
2009-01-01
Abstract:Objective To verify the accuracy of vector flow mapping (VFM) in fluid quantification in vitro. Methods An experimental hydrodynamic model was established to provide stable flow at specified speed in vitro. The VFM measurement was performed to quantify the flow velocity and rate. The results of VFM quantification were compared with the actual flow rate to verify the accuracy of VFM method. ResultThe flow rate obtained by VFM measurement was significantly correlated with the genuine flow rate (r=0.997,b=0.949,P=0.000). The within variability was 1.3%. Conclusion The VFM is an accurate method with high reproducibility in flow quantification.