Volume flow estimation using time-domain ultrasonography:in vivo study

张韵华,徐智章,刘利民,冯岚,徐佰成
DOI: https://doi.org/10.3760/j.issn:1004-4477.1999.05.017
1999-01-01
Abstract:Objective On the base of in vitro experiment,accuracy of color velocity imaging quantification (CVIQ) were further studied in vivo on pulsative flow,flexible artery.Methods Four dogs were cannulated (two were common carotid artery to common carotid artery,two were common carotid artery to internal jugular vein).Volume flow estimation were excutated simultaneously using CVIQ and bubble method under different volume flow level.In addition,CVIQ and timed-collection volume flow estimation were taken simultaneously in one dog.Results Under 5 group (mean volume flow from 60~370 ml/min),totle 65 times estimation of CVIQ and bubble method.two methods were in linear with regression slope 1.0125 and correlation coefficient 0.975,P=0.0001.Compared to timed-collection,the relative error of CVIQ were-13.3%~20.6%,mean error 4.7% under mean volume flow 465 ml/min(10 times comparison).Coefficient of variation (CV) of CVIQ were somewhat greater than timed-collection and bubble method's.Conclusions Noninvasive way of accurate volume flow estimation jusitifies CVIQ as a practicable way of volume flow estimation.