The Clinical Evaluation for the Distorted Rate of Digital- Panoramic Radiograph in Dental Implant

WANG Hu,ZHANG Jing,NIU Yuming
DOI: https://doi.org/10.3969/j.issn.1007-3957.2006.02.007
2006-01-01
Abstract:Objective:The purpose of this article is to evaluate the guiding signaficience of the digital-panoramic radiograph in the clinical practice. Methods:56 patients with a total of 162 dental implants were selected for this study. The metal balls with standard size which be put in the template were used at preoperation. After acquairing the digital-panoramic radiograph,we applied the Cliniview5.1 analysis software to measure diameter of the metal balls which lied in various sites in perpendicular dimention three times. Then we acquaired the means. Throught study the change of diameter of the metal balls in the digital-panoramic radiograph,,we can quantify accurate bone height in different areas of the maxillary and mandibular dental arch. Results: The discrepancies between measure results and standard diameter were very minute. The range of distortion is 0.2%~2%,but the measurements of different areas have significant differences (P<0.05):1) the mandibular anterior areas and mandibular premolar areas; 2)the mandibular anterior areas and mandibular molar areas; 3)the mandibular premolar areas and the maxillary molar areas; 4)the maxillary and mandibular molar areas. Conclusions Before implanted operation,we can employ the technique of digital-panoramic to evaluate the bone height of preaparing implant areas,and it may help us achieve accomplishment. Though the measurements of different areas have discrepancies,it have no significant effects on accuracy .
What problem does this paper attempt to address?