Experiemtal Investigation of the Bending Fatigue Strength for Balloon Expandable Coronary Stents

Feng Haiquan,He Pingji,Hu Zhiyong,Li Zhiguo,Han Qingsong
DOI: https://doi.org/10.3969/j.issn.1006-2343.2013.02.010
2013-01-01
Abstract:To reveal the fracture reason in coronary stent implantation,the stent elastoplastic deformation as well as its fatigue life is investigated under the pulse induced fluctuating load.On the basis of F2477-07 standard from American Society for Testing and Materials(ASTM),the stent bending fatigue life test method is proposed during coupled expansion deformation by Roark stress-strain method.The key technology concerned with modeling the coupled stent system is studied with more attention to vascular model,loading definition,physiological conditions,test method and performance assessment.Subsequently the accelerated fatigue life test method under the fluctuating implse load is put forward.Tesing the bending fatigue life of six different stents,the fracture reason is analysed to reveal the effect of geometrical demension,mechanical performance and material property on the stent bending fatigue life.
What problem does this paper attempt to address?