Determination Of Diethylstilbestrol In Human Plasma With Measurement Uncertainty Estimation By Liquid Chromatography-Tandem Mass Spectrometry
Yi Liu,Na Li,Liping Ma,Lin Huang,Shan Jing,Libo Zhao,Wanyu Feng
DOI: https://doi.org/10.1080/10826076.2012.745140
2014-01-01
Abstract:A rapid and sensitive liquid chromatography/tandem mass spectrometry (LC-MS/MS) using an electrospray ionization (ESI) method was developed and validated to determine the concentration of diethylstilbestrol in human plasma. Samples were extracted with methyl tert-butyl ether after alkalified by sodium tetraborate solution, using 3,5-dihydroxy-4-isopropyl toluylene as the internal standard (IS). Separation was performed on a Restek C-18 column (2.1mmx150mm, 5.0 mu m) guarded by a Hypersil Gold pre-column (2.1mmx10mm, 3 mu m). The mobile phase consisted of acetonitrile-water (85:15, containing 0.2mmol/L ammonium acetate) at a flow rate of 0.3mL/min. Detection was performed using negative multiple reaction monitoring (MRM) mode. The mass transitions monitored were [M-H](-)m/z267.2237.2 and [M-H](-)m/z253.2211.0 for diethylstilbestrol and IS, respectively. The method was fully validated using total error theory, which is based on -expectation tolerance intervals and include trueness and intermediate precision. The method was found to be accurate over a concentration range of 0.1-50ng/mL. The main advantages of this method were the high sensitivity (a lower limit of quantification, LLOQ, of 0.1ng/mL) and the short analysis time (total run time 5min per sample). The measurement uncertainty based on -expectation tolerance intervals was assessed at each concentration level of the validation standards. The method is specific, sensitive, and accuracy, and it is suitable for plasma concentration determination of diethylstilbestrol.