Liquid Chromatography Tandem Mass Spectrometry Assay to Determine the Pharmacokinetics of Aildenafil in Human Plasma

Jiang Wang,Yao Jiang,Yingwu Wang,Xia Zhao,Yimin Cui,Jingkai Gu
DOI: https://doi.org/10.1016/j.jpba.2007.01.017
IF: 3.571
2007-01-01
Journal of Pharmaceutical and Biomedical Analysis
Abstract:A simple, sensitive and specific liquid chromatography/tandem mass spectrometry method for the quantitation of aildenafil, a new phosphodiesterase V inhibitor, in human plasma is presented. The analyte and internal standard, sildenafil, were extracted by a one-step liquid–liquid extraction in alkaline conditions and separated on a C18 colunm using ammonia:10mM ammonium acetate buffer:methanol (0.1:15:85, v/v/v) as the mobile phase. The detection by an API 4000 triple quadrupole mass spectrometer in multiple-reaction monitoring mode was completed within 2.5min. The calibration curve exhibited a linear dynamic range of 0.05–100ng/ml with a 10pg/ml limit of detection. The intra- and inter-day precisions measured as relative standard deviation were within 8.04% and 5.72%, respectively. This method has been used in a pharmacokinetic study of aildenafil in healthy male volunteers each given an oral administration of one of the three dosages.
What problem does this paper attempt to address?