Biomechanical research of a new injectable prosthetic nucleus pulposus

姬勇,邹德威,吴继功,庹新林,王晓工,王婧,周雪峰,马华松
DOI: https://doi.org/10.3969/j.issn.1004-406X.2009.05.14
2009-01-01
Abstract:Objective:To test the biomechanical features of a new injectable prosthetic nucleus pulposus so as to provide data for clinical practice.Method:The biomechanical test was carried out on prosthetic nucleus pulposus prepared by filling with polyurethane and silicone rubber to explore compression modulus,static stiffness,dynamic stiffness and the yielding limit.The relevant data were recorded and analyzed.Result:When increasing the preload to 150N,250N,350N,500N,1000N with a speed of 10N/s,the stress and strain of prosthetic nucleus pulposus showed a non-linear relationship with the elastic modulus of approximate 2.5MPa.During the process of increasing the preload to 1000N with a speed of 10N/s,the static stiffness increased gradually and demonstrated non-linear feature.During the process of increasing the preload to 400N with a speed of 10N/s,the dynamic stiffness in range of 0.5-3Hz laged behind the stress of the phase angle at about 13 degrees.Dynamic stiffness was about 3.5 times of static stiffness in the frequency of 3Hz.During the process of increasing the preload from 500N to 2000N with a speed of 10N/s,the filler showed no plastic deformation,while at the preload of 1200N,1 prosthetic nucleus pulposus broken.Conclusion:New injectable prosthetic nucleus has good flexibility and toughness,and it is close to the physiological characteristics of real nucleus pulposus,which may be used in animal research,however this material is something to be improved.